Appendix I: Calibration

Calibration of the supply needs to be performed on each of the different sections, positive 5V, positive 15V, negative 15V, positive variable, negative variable. Within each section, both the current and voltage inputs to the A/D are adjusted for maximum value and for zero offset. In addition the variable supplies have to be adjusted for their maximum output voltage prior to calibration of the A/D inputs.

Positive 5V Section

Please consult Figure 12 as you follow the instructions for calibrating the fixed positive 5V section. Fixed Section Calibration Photograph
Figure 12: Fixed Sections: +5V and ±15V
Resistor and VR labels were lost during translation; please contact the authors if you require these details.

To calibrate the 5V voltage monitoring section the following steps are taken:

  1. Connect DVM from pin 9 of A/D to ground.
  2. Remove +5V fuse.
  3. Adjust VR3 for 0.00V on DVM.
  4. Reinsert +5V fuse.
  5. Adjust VR4 for 4.50V on DVM.
  6. Repeat steps 2 to 5 until adjustments are reached for both steps 3 and 5.

To calibrate the 5V current monitoring section the following steps are taken:

  1. Connect DVM from pin 10 of A/D to ground.
  2. Remove +5V fuse.
  3. Install jumper across RSENS.
  4. Adjust VR1 for 0.00V on DVM.
  5. Remove jumper across RSENS.
  6. Reinstall +5V fuse.
  7. Load supply 5W, 5W on output terminals.
  8. Adjust VR2 for 2.50V.
  9. Repeat steps 2 to 8 till adjustments are reached for both steps 4 and 8.

15V Section

Please consult Figure 12 as you follow the instructions for calibrating the fixed positive 15V section.

Positive

To calibrate the +15V voltage monitoring section the following steps are taken:

  1. Connect DVM from pin 9 of A/D to ground.
  2. Remove +15V fuse.
  3. Adjust VR7 for 0.00V on DVM.
  4. Reinsert +15V fuse.
  5. Adjust VR8 for 4.50V on DVM.
  6. Repeat steps 2 to 5 till adjustments are reached for both steps 3 and 5.

To calibrate the 15V current monitoring section the following steps are taken:

  1. Connect DVM from pin 10 of A/D to ground.
  2. Remove +15V fuse.
  3. Install jumper across RSENS.
  4. Adjust VR5 for 0.00V on DVM.
  5. Remove jumper across RSENS.
  6. Reinstall +15V fuse.
  7. Load supply 15W 5W resistor on output terminals.
  8. Adjust VR6 for 2.50V.
  9. Repeat steps 2 to 8 till adjustments are reached for both steps 4 and 8.

Negative

Please consult Figure 12 as you follow the instructions for calibrating the fixed negative 15V section.

To calibrate the -15V voltage monitoring section the following steps are taken:

  1. Connect DVM from pin 9 of A/D to ground.
  2. Remove -15V fuse.
  3. Adjust VR11 for 0.00V on DVM.
  4. Reinsert -15V fuse.
  5. Adjust VR12 for 4.50Von DVM.
  6. Repeat steps 2 to 5 till adjustments are reached for both steps 3 and 5.

To calibrate the -15V current monitoring section the following steps are taken:

  1. Connect DVM from pin 10 of A/D to ground.
  2. Remove -15V fuse.
  3. Install jumper across RSENS.
  4. Adjust VR9 for 0.00V on DVM.
  5. Remove jumper across RSENS.
  6. Reinstall -15V fuse.
  7. Load supply 15W 5W resistor on output terminals.
  8. Adjust VR10 for 2.50V.
  9. Repeat steps 2 to 8 till adjustments are reached for both steps 4 and 8.

Variable Section

Positive

Please consult Figure 13 as you follow the instructions for calibrating the variable positive supply. Positive Variable Section Calibration Photograph
Figure 13: Positive Variable Supply

The first steps in the adjustments for the variable supplies is to insure that the Output voltage maximum is set to the correct value. To do this the following steps are taken:

  1. Set the input word to maximum value.
  2. Connect DVM from the output to ground.
  3. Adjust the trim pot for the R4 till the DVM reads 36.00V.

To calibrate the positive variable voltage monitoring section the following steps are taken:

  1. Connect DVM from pin 9 of A/D to ground.
  2. Set voltage to zero.
  3. Adjust VR3 for 0.00V on DVM.
  4. Set voltage to 36.00V.
  5. Adjust VR4 for 5.00V on DVM.
  6. Repeat steps 2 to 5 till adjustments are reached for both steps 3 and 5.

To calibrate the +Variable current monitoring section the following steps are taken:

  1. Connect DVM from pin 10 of A/D to ground.
  2. Set voltage to zero.
  3. Install jumper across RSENS.
  4. Adjust VR1 for 0.00V on DVM.
  5. Remove jumper across RSENS.
  6. Set voltage to 17.50V.
  7. Load supply 17.5W 5W resistor on output terminals.
  8. Adjust VR2 for 2.50V.
  9. Repeat steps 2 to 8 till adjustments are reached for both steps 4 and 8.

Negative

Please consult Figure 14 as you follow the instructions for calibrating the variable negative supply. Negative Variable Section Calibration Photograph
Figure 14: Negative Variable Supply

The first steps in the adjustments for the variable supplies is to insure that the output voltage maximum is set to the correct value. To do this the following steps are taken:

  1. Set the input word to maximum value.
  2. Connect DVM from the output to ground.
  3. Adjust the trim pot for the R4 till the DVM reads -36.00V.

To calibrate the negative variable voltage monitoring section the following steps are taken:

  1. Connect DVM from pin 9 of A/D to ground.
  2. Set voltage to zero.
  3. Adjust VR3 for 0.00V on DVM.
  4. Set voltage to -36.00V.
  5. Adjust VR4 for 5.00V on DVM.
  6. Repeat steps 2 to 5 till adjustments are reached for both steps 3 and 5.

To calibrate the negative variable current monitoring section the following steps are taken:

  1. Connect DVM from pin 10 of A/D to ground.
  2. Set voltage to zero.
  3. Install jumper across RSENS.
  4. Adjust VR1 for 0.00 V on DVM.
  5. Remove jumper across RSENS.
  6. Set voltage to -17.50 V.
  7. Load supply 17.5W 5W resistor on output terminals.
  8. Adjust VR2 for 2.50V.
  9. Repeat steps 2 to 8 till adjustments are reached for both steps 4 and 8.